  • Linear range: 0 to 20 ppm (mg/L) as Cl2
  • Linearity per ISO 15839: 0 - 10 ppm:2%; 0 - 20 ppm: 3%
  • Response time following step change in concentration: 90% of final value within 7 minutes at 25°C
  • Drift: ~1.5ppm in clean water and constant temperature; < 0.05 ppm over 2 weeks
  • Detection limit (per ISO 15839): 0.02 ppm (mg/L) in clean water at room temperature

Description

The Rosemount Analytical TCL Total Chlorine Analyzer is intended for the determination of total chlorine in water, including the determination of chlorine in seawater. The Rosemount Analytical TCL system consists of a sample conditioning unit, a sensor and either the 1056 Dual Input Analyzer (1056-24) or 56 Dual Input Analyzer (56-24). With no metal wetted parts this Rosemount Analytical TCL system is ideal for seawater. The sample flow rate is approximately a low 15mL/minute resulting in very little waste. The reagent-based system measures true total chlorine.

The Rosemount Analytical TCL Total Chlorine Analyzer is a sample conditioning system that permits a single sensor to measure total chlorine in water. The sample conditioning system continuously injects a solution of acetic acid (vinegar) and potassium iodide into the sample. The acid lowers the pH to between 3.5 and 4.5 and allows total chlorine in the sample to quantitatively react with the potassium iodide concentration, and the analyzer displays the total oxidant concentration in ppm as Cl 2.

Accessories

Required Components

These additional components are required for the operation of your product. We strongly advise you to purchase these components unless you already own them or own a suitable alternative.

Mounting Bracket

Sensors/Probes

Solutions

Accessories

Please consider these optional accessories.

Applications

This product can be used in the following applications: