Starting at $29,995.00
  • Usually ships in 4-6 weeks

Overview

Features

  • 0.008% RD, 0.007% FS, or 0.01% FS precision
  • 0.01% RD, 0.01% FS, or 0.02% FS accuracy
  • <0.005% FS control stability, typically 0.003% FS
  • <30s control response time
  • Gauge, absolute pressure type
  • 1 internal pressure control module
    • Max range: 0 to 15,000 psi
    • Min range: -15 to 3,000 psi
  • 1 external pressure control module
    • Min range: -15 to 1,000 psi
  • Fixed or auto range switching mode
  • Built-in hydraulic pump, no external pressure source required
  • Emulation mode

What's in the Box

  • (1) ADT 793 pressure controller
  • (1) ISO17025 accredited calibration certificate
  • (2) Green terminal plug for switch test
  • (10) O-ring 3.5*1.5 for sealing pressure module
  • (1) Drain switch valve assembly set (switch valve + 0.5m x 6mm tube)
  • (1) Accumulator inflation assembly set (adapter + 1.5m tube for gas inflation) (oil media only)

Description

The Additel ADT 793 Pressure Controller is ideal for users requiring automated production, testing, and calibration. The ADT 793 accepts one ADT151 dual-range pressure module in addition to a barometric pressure module, allowing users to quickly cover a wide range of pressures.

An ADT151 pressure module within the range of GP15KM to CP3KM must be installed.

The ADT151 pressure modules can be installed or replaced on the ADT 793 controller within 30 seconds or less. As the cabin door opens, the controller will automatically release pressure, providing the safe removal and installation of the ADT151 modules. Select between 15,000 psi, 10,000 psi, 6,000 psi, 5,000 psi, 3,600 psi, or 3,000 psi with a variety of accuracies to meet application requirements.

The ADT 793 adopts professional control technology to effectively improve control rate and stability: control response time (typical) ≤30 seconds, and control stability (typical) ≤±(0.003 to 0.005)% FS. The controller also supports switching between internal and external liquid storage tanks, allowing users to quickly connect the large capacity external liquid reservoir.