Additel ADT 773 Pressure Controller
Pressure ranges from -380 to 400 inH2O.
Overview
Features
- Precision:
- 0.015% FS (DP2 to DP5 modules)
- 0.025% FS (DP10 to DP1K modules)
- Accuracy:
- 0.05% FS (DP2 to DP5 modules)
- 0.02% FS (DP10 to DP1K modules)
- Pressure control range of internal module:
- Max: -400 to 400 inH2O
- Min: -1 to 1 inH2O
- Control stability of 0.003% FS
- Option to install barometric pressure module
- <10s control response time
- Low differential pressure type
- 2 interchangeable pressure module bays
- Fixed or auto range switching mode
- Built-in differential pressure pump
- 7in color touch screen display
- Emulation mode
What's in the Box
- (1) ADT 773 pressure controller
- (1) AC power cord (10A 250V)
- (1) ISO17025 accredited calibration certificate
- (2) Green terminal plug for switch detection
- (10) O-ring 3.5*1.5 for ADT151
- (2) Festo plug 6mm for sealing
- (3) 0.3m silicon tube, 120 kPa max
- (2) 1.5m polyurethane tube, 700 kPa max
- (2) ADT100-BARB-FESTO
Description
The Additel ADT 773 Pressure Controller is optimized for low-pressure work, utilizing Additel's low-pressure pump technology in combination with the control capabilities of the handheld model ADT760. The ADT 773 is ideal for users requiring automated production, testing, and calibration. The ADT 773 can be outfitted with two controlling ADT151 modules and one reference barometric module to cover a wide pressure range.
The ADT151 pressure modules can be installed or replaced within 30 seconds or less. As the cabin door opens, the controller will automatically release pressure, providing the safe removal and installation of the ADT151 modules.
At least one ADT151 Digital Pressure Module within the range of DP30 to DP400 must be installed in the high-side module bay. The low bay can be left empty, but when a low-pressure module is installed, the range must be lower than the installed high-pressure module.
The ADT 773 has an integrated internal low-pressure pump that gives users a turnkey solution for pressure calibration work. The controller provides valve control for up to 3 external valves. One of the channels is designated for controlling a contamination prevention system (CPS), which helps to prevent media contamination. The remaining two channels can be used to control external vacuum pumps or external isolation valves for greater flexibility and compatibility for individual situations.
Documents
Accessories
Required Components
These additional components are required for the operation of your product. We strongly advise you to purchase these components unless you already own them or own a suitable alternative.
Accessories
Please consider these optional accessories.
Need Help? Call a Calibration engineer at 1-800-884-4967