E+E EE610 Low Differential Pressure Sensor
Designed for highly accurate measurement in applications like medical, pharmaceutical and clean rooms; selectable ranges to 0.4inH2O.
Overview
Features
- Selectable pressure range:
- ±0.1, ±0.2 and ±0.4inH2O or
- 0 to 0.4inH2O
- ±0.5% FS accuracy
- <0.002inH2O long-term stability
- ±28inH2O overload limit
- -4 to 140°F (-20 to 60°C) temperature range
- Digital and analog output
- IP65/NEMA 4 protection
- Large graphic display
- Installation-friendly enclosure
Description
The E+E EE610 low differential pressure sensor can be employed in air or in non-flammable and non-aggressive gases. For analog output versions, the EE610 offers measuring ranges selectable with DIP switches. The piezo-resistive, no flow-through pressure sensing element eliminates the risk of cross-contamination and offers long-term stability.
The measured data is available on the analog voltage and current outputs, or on the RS485 interface with Modbus RTU or BACnet MS/TP protocol. The IP65/NEMA 4 enclosure facilitates easy mounting and minimizes installation costs. External mounting holes allow installation with a closed cover, protecting the electronics against construction site damage and pollution.
The setup and adjustment can easily be performed with DIP switches and push buttons on the electronics board (analog output version), or with an optional adapter and the EE-PCS configuration software (RS485 interface version). The display includes bidirectional or unidirectional pressure range, output signal, response time, display units and backlight. A zero and span point adjustment can be performed with push buttons on the electronics board.
Documents
Documents
- E+E EE610 Low Differential Pressure Sensor Datasheet (1.7 MB)
- E+E EE610 Low Differential Pressure Sensor Declaration of Conformity (64 KB)
- E+E EE610 Low Differential Pressure Sensor Quick Guide (1.1 MB)
- E+E EE610 Low Differential Pressure Sensor BACnet Protocol Implementation (255 KB)
- E+E EE610 Low Differential Pressure Sensor Manual (2.2 MB)
- All possible part numbers
Need Help? Call a Calibration engineer at 1-800-884-4967