Testo 6721 Dew Point Switch
Humidity monitoring in compressed air down to -30°C.
Overview
Features
- Detects moisture in compressed air systems
- Measurement range -30 to 30°C (-22 to 86°F)
- Optional alarm box available
- Easy configuration with P2A software
- Uses proven Testo humidity sensor
- Mini DIN interface
Description
The Testo 6721 dew point switch measures the temperature pressure dew point (tpd) coming from refrigeration dryers or downstream in pneumatic machines and processes from -30 to 30°C (-22 to 86°F) up to 300 psi (20 bar). Moisture in compressed air lines can lead to expensive damage: icing, corrosion, water hammer, sticky and clogged materials. The Testo 6721 dew point switch can save time, money and system downtime by monitoring moisture and avoiding damage.
The Testo 6721 dew point switch uses Testo's proven RH sensor design for accurate temperature pressure dew point measurements. It's easily installed in dryer and compressed air lines via its 1/2" NPT threaded fitting.
The Testo 6721 dew point switch can be configured using the optional Testo P2A software via USB cable and the data port on the 6721 switch. The software can easily program, adjust and review the 6721 operating history while connected to a laptop or PC. The user can even save a "master" program file that provides easy "copy and paste" program replication for OEM requirements. An optional alarm box (0554 6722) provides visual confirmation of both pre and alarm conditions.
Testo 6721 Dew Point Switch Specifications
Feature | Testo 6721 |
---|---|
Measuring range | -30 to 30°C (-22 to 86°F) |
Accuracy | ±4 K (-30 to -20°C)(-22 to -4°F) ±3 K (-20 to -10°C)(-4 to 14°F) ±2 K (-10 to 0°C)(14 to 32°F) ±1 K (>0°C)(>32°F) |
Sensor | Testo humidity sensor (with special trace humidity adjustment) and NTC temperature sensor |
Process pressure | Max. 20 bar (abs)(290 psi) |
Process temperature | 0 to 50°C (32 to 122°F) |
Outputs | 2x switches potential-free max. 30 V/0.5 A |
Interface | Mini-DIN interface (serial) for parameterization / adjustment / analysis via P2A software |
Voltage supply | 20 to 30 VAC/VDC |
Documents
Accessories
Power Supplies
Software
Accessories
Please consider these optional accessories.
Need Help? Call a Temperature engineer at 1-800-884-4967
We're open Mo-Th 8am to 5:30pm. Fr 8am to 5pm ET