MBW Calibration 473 Dew Point Hygrometer
Highly precise chilled mirror hygrometer for dew points between -30 and 99°C.
Overview
Features
- Dew point range:
- RP2: Direct insertion, -20 to 70°C (-4 to 158°F)
- SH2: Flow-through, -20 to 70°C (-4 to 158°F)
- SHX: Flow-through, -30 to 99 °C (-22 to 210.2°F)
- Temperature range:
- RP2 & SH2: -20 to 80°C (-4 to 176°F)
- SHX: -50 to 125°C (-58 to 257°F)
- Accuracy Dew point/Temperature
- RP2: ≤ ±0.15°C/≤ ±0.07°C
- SH2: ≤ ±0.1°C/≤ ±0.07°C
- SHX: ≤ ±0.1°C/≤ ±0.07°C
- Power Supply: 100-24VAC
- Internal Barometric Pressure Sensor, 700 to 1200mbar (optional)
- 0.1% or 0.01% accuracy, 700 to 1200mbar
- Two user programmable analog outputs
- -10 to 10V
- 4 to 20mA
- RS-232 digital I/O
- Suitable for use as a transfer standard
- 5 points FP/DP, 3 points temp. factory calibration certificate
- Pelican 1550 foam lined transport case
What's in the Box
- Measuring head
- 2m (6ft) cable
- External temperature probe
- Operation and maintenance manual
- Calibration certificate
- Transport case
Description
The MBW Calibration 473 dew point hygrometer features cable mounted measuring heads that allow the user to measure directly within the application environment, and includes a temperature probe as standard. The 473 is able to calculate, display and output relative humidity.
MBW chilled mirror hygrometers provide precise, stable and repeatable measurement results. The 473 utilizes high quality materials and precision engineering have a service life of more than 15 years. The platinum resistance thermometer (PRT) element that is embedded within the mirror ensures measurement stability over a long period of time.
MBW's ForceFrost™ function over-cools the mirror to force the condensed layer to the solid phase when the temperature falls below a user defined limit. This eliminates the uncertainty of whether dew or frost point is measured.
Three different measuring heads with 2m connecting cables are available.
SH2 Measuring Head
Flow-through dew point measuring head with a 2-stage Peltier element for mirror temperature control. Includes a calibrated temperature probe for connecting to either the measuring head or the back panel of the 473. The included variable speed fan pulls a consistent airflow across the mirror. With the fan removed, the user can also connect the SH2 to applications using tubing and standard 6mm or 1/4in fittings.
RP2 Measuring Head
A 2-stage Peltier element in a compact probe that includes a connection for temperature measurement. It is supplied with a calibrated head mounted temperature probe and an extension cable to enable optimum placement for your application.
SHX Measuring Head
Chilled mirror dew point head designed for high temperature applications. It incorporates a sample fan with controllable speed to maintain stable and precise flow within the recommended range of 0.3 to 1lpm at atmospheric pressure. A high temperature motor is combined with an aluminum impeller/cowl to maintain a reliable and stable sample flow. The flow rate can be changed through serial commands that are entered via the front panel of the 473.
Documents
Accessories
Calibration
Upgrades
Cables
Sensors/Probes
Need Help? Call a Calibration engineer at 1-800-884-4967
We're open Mo-Th 8am to 5:30pm. Fr 8am to 5pm ET