Starting at $1,045.47
  • Usually ships in 3-6 weeks

Overview

Features

  • 0 to 14 pH measuring range
  • Chemically resistant KYNAR
  • Rugged, quick change, quick clean
  • Max temp: 266°F/130°C at 40 psig/276 KPa
  • Max pressure: : 150 psig/1035 KPa at 158°F/70°C
  • Model RBI-546:
    • 3/4in MPT threaded nose
    • Submerged; 3/4in MNPT threaded top
  • Model RBI-547:
    • 1in MPT process fitting or, 1¼ and 1½in ball valve assemblies
    • 1in MNPT process fitting, into 1in FNPT coupling or ¾in submersion extension pipe

Description

The Rosemount Analytical RBI pH Sensor is designed for extended sensor life when used in applications with sulfides which are typically found in scrubbers. The Model RB is made with chemically resistant KYNAR. Optional installation methods allows standardization of entire plant with a minimum of sensor part numbers.

They're built with a Kynar body and is designed in two configurations. The 546 sensor is housed in a fully encapsulated body for use in pipe tees or submersed in a tank. The 547 is the ball valve retractable version; the sensor body can be replaced and the titanium sensor sheath assembly (sold separately) can be reused. Both sensor types are built with a high performance J glass, notched tip with a Teflon junction, and a PT100 RTD. The integral cable is prepared with tinned leads for easy wiring.

Accessories

Required Components

These additional components are required for the operation of your product. We strongly advise you to purchase these components unless you already own them or own a suitable alternative.

Cables

Solutions

Accessories

Please consider these optional accessories.

Applications

This product can be used in the following applications:

Here's how some of our customers use this product...

Monitoring the pH of a Wet Scrubber

The Background: Our customer is an industrial facility who’s process creates nitrogen oxide as a byproduct. Nitrogen oxide is a powerful pollutant that needs to be removed from the plant’s emissions before they are released to the atmosphere. It is removed using a wet scrubber. The customer needs an effective means of monitoring the pH of the wet scrubber to ensure the proper concentration of the chemicals used to remove the nitrogen oxide.

The Problem: The wet scrubber uses sodium hydroxide and sodium hydrosulfide to remove the nitrogen oxide. Both of these compounds are powerful alkalis meaning the pH can be as high as 12. A pH that high cannot be accurately measured with many pH electrodes. The pH is also high enough to damage electrodes.

The Solution: We recommended the Rosemount Analytical RBI pH Sensor which is designed for challenging environments. The RBI has a measuring range up to 14 pH and has a chemically resistant KYNAR body. The RBI includes has 5 junctions to resist poisoning extend sensor life.