Starting at $17,935.00
  • Usually ships in 2-3 weeks

Overview

Features

  • Accurate to ±0.04°C using the external reference sensor
  • Excellent stability: 0.01°C
  • Wide temperature range
    • RTC-158: -22 to 155°C (-8 to 311°F)
    • RTC-250: 28 to 250°C (82 to 482°F)
  • Unique active dual-zone block for improved temperature homogeneity
  • Dynamic Load Compensation for perfect temperature uniformity in the insert
  • Display indicator for temperature uniformity
  • Intelligent reference sensors hold the calibration data (coefficients) of the reference sensor
  • USB communication
  • High speed heating and cooling times
  • Best performing dry-block with regard to the EURAMET/cg-13/v.01 guideline for testing of dry-blocks

Description

The Ametek Jofra RTC-158 and RTC-250 are advanced reference temperature calibrators. Combining dry-block and liquid bath technology, the RTC-158 and RTC-250 are designed to calibrate oddly sized and shaped sensors or multiple sensors at once. Both calibrators offer a host of innovative features along with AMETEK's long history of providing reliable, low-drift calibration instruments.

The Ameteck Jofra RTC-158 has a temperature range of -22° to 155°C (-8° to 311°F), while the RTC-250 has a temperature range of 28° to 250°C (82° to 482°F). Both units boast outstanding accuracy and stability while featuring Ametek's Dynamic Load Compensation (DLC) system that provides perfect temperature uniformity in the insert no matter the number or size of the sensors immersed. The calibrators also offer unique intelligent sensors for plug ‘n' play connectivity, an easy-to-read full-color 5.7-inch VGA display, and a USB connection for improved communications and software downloads.

The Ametek Jofra RTC-158 and RTC-250 reference temperature calibrators are available in three configurations: RTC-A reference temperature calibrator, RTC-B reference temperature calibrator with input for reference sensor, DLC sensor and sensors-under-test, and RTC-C reference temperature calibrator with input for reference sensor and DLC sensor. The RTC-158 and RTC-250 can be used without an external reference sensor. However, users will obtain better accuracies and realize the calibrator's full potential by utilizing the STS-200 reference sensor. Both the STS-200 and DLC sensors contain all individual calibration data regarding the sensor, allowing users to change the reference sensor and be up and running immediately.

The RTC-158 and RTC-250 feature a rugged, lightweight design and multi-hole inserts that cover most commonly used sensor sizes. A unique design allows the units to calibrate threaded sensors and sensors with connection heads. The 160 mm deep well has a diameter of 63.5 mm, which is twice the size of any other dry-block calibrator. Either calibrator can be used both as a dry-block or liquid calibration bath with magnetic stirrer.

The Ametek Jofra RTC-158 and RTC-250 (models A, B and C) ship complete with the following:

  • Mains power cable
  • Traceable calibration certificate (temperature performance)
  • Tool for insertion tubes
  • JOFRACAL software
  • AMETRIM software
  • USB cable
  • Set of rubber cones for insulation plugs
  • Manuals

Model B contains the following extra items:

  • Test cables (2 x red, 2 x black)
  • Traceable certificate (input performance for reference sensor and DLC sensor)
  • Traceable certificate (input performance for sensor-under-test inputs)

Model C contains the following extra items:

  • Traceable certificate (input performance for reference sensor and DLC sensor)

Ametek Jofra RTC-158 and RTC-250 Temperature Calibrators Specifications:

Parameter RTC-158 RTC-250
Temperature Range @ 0°C ambient -37 to 155°C 5 to 250°C
@ 23°C ambient -22 to 155°C 28 to 250°C
@ 40°C ambient -9 to 155°C 45 to 250°C
Accuracy With external ref. sensor (models B&C) ±0.04°C ±0.07°C
With internal ref. sensor ±0.18°C ±0.28°C
Stability ±0.01°C ±0.02°C
Radial Homogeneity Block to 0.03°C to 0.05°C
Bath to 0.15°C to 0.15°C
Resolution All temperatures 1°, 0.1°, 0.01°, or 0.001°
Heating Time -22 to 23°C 9 minutes
23 to 100°C 23 minutes
100 to 155°C 23 minutes
28 to 100°C 3 minutes
50 to 100°C 2 minutes
100 to 250°C 9 minutes
Cooling Time 155 to 100°C 9 minutes
100 to 23°C 24 minutes
23 to 0°C 15 minutes
0 to -15°C 21 minutes
250 to 100°C 24 minutes
100 to 50°C 24 minutes
100 to 28°C 65 minutes
Time to Stability 15 minutes
Immersion Depth With insulation plug 180 mm
Bath version 150 mm
Input Specifications RTD reference input See data sheet for complete specifications
DLC sensor input See data sheet for complete specifications
RTD sensor under test See data sheet for complete specifications
Thermocouple input See data sheet for complete specifications
Transmitter Supply Output voltage 24 VDC ±10%
Output current Maximum 28 mA
Transmitter Input mA (B Model) Range 0 to 24 mA
Accuracy (12 months) ±(0.005% rdg + 0.010 F.S.)
Voltage Input VDC (B Model) Range 0 to 12 VDC
Accuracy (12 months) ±(0.005% rdg + 0.010 F.S.)
Switch Input (B Model) Test voltage Maximum 5 VDC
Test current Maximum 2.5 mA
Communications Interface Serial data USB 2.0 device port
Serial data USB 2.0 host double port (for future expansion)
LAN Ethernet MAC 10/100 Base-T (for future expansion)
SD SD slot (for future expansion)
Inserts Outer diameter 63.5 mm
Length 160 mm