Changing the type of input signal on the PHA recorder
If you have a question not answered here, please contact us.
This recorder is a multi-input device which permits the input for any channel to be changed to thermocouple, RTD or DC voltage input. Follow the procedure described below to change the input type.
Step 1:
Turn off the power to the unit.
-
Step 2:
Open the front door and remove the main unit in the manner shown in the drawings below.
-
Step 2-1
Loosen the lock screws on the left and right side and remove them.
Step 2-2
Grip the paper feed frame and pull it firmly towards you. This detaches it from the main unit.
Step 2-3
Change the settings of the pins for individual channels on the internal printed circuit board.
*Use pliers to remove and insert pins
Step 2-4
After completing the hardware change, push the main unit back into its original place and screw back into position.
Step 2-5
Change the input terminal wiring to make it correspond to the new input signal type. For DC current input, provide the input terminals with shunt resistors.
Step 2-6
Refer to Section 7.8 in the manual and carry out the programming changes for each channel that has been modified.
Questions?
Feel free to give us a call at 1-800-884-4967